ProCure Manual Firmware Update

ProCure Manual Firmware Update

Please follow the step-by-step guide below to manually updated your ProCure firmware:

1

Turn off Pro Cure and unplug it from power

  • Toggle the rear power switch to the off position

  • Unplug the power cable from the power plug port

  • Note: the Pro Cure must be disconnected from power to successfully update

2

Connect Pro Cure to your computer with a USB-A to USB-B cable

● This must be done only after the machine is disconnected from the power

3

Download these two files to your computer

(Right click to save the files below)

version.txt (4 Bytes)

procure.bin(400 KB)

4

Format the Drive using Disk Utility

You will format the drive via Disk Utility. When opening Disk Utility on your Mac, you should see the ProCure show up as a USB drive on the left side.​ Select the drive and click "Erase". You should see a window pop up, please ensure the Format Selected is ExFAT as this is the format compatible with both Mac and Windows​. Once selected, click erase. This will erase and format the drive.

Once you are done with the format, open a new finder window. Drag and drop both procure.bin and version.txt files onto the ProCure's USB drive.​​​​ ​​

Once you've copied both files onto the drive, you will eject the drive. Unplug the USB cable from the ProCure

5

Power on the Pro-Cure

Plug in the power cord
Turn the Pro-Cure power switch to the on position