Before You Begin
- Always wear nitrile gloves when handling items that come into contact with uncured resin.
Manual Debris Removal Procedure
- Select Settings > Manual Debris Removal to open the tab.

- Set the required exposure time using the + and – icons.
- Select Expose. The entire material tray area is exposed.
- Remove debris from the material tray by lifting the thin film of the exposed material.
- Proceed following the step-by-step for filtering material described on the Einstein / Pro XL: Maintain Material page, if needed.